MAUDE Adverse Event Report: BECTON DICKINSON & CO. (SPARKS) BD BACTEC¿ FOS¿ (FASTIDIOUS ORGANISM SUPPLEMENT) KIT; SYSTEM, BLOOD CULTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON & CO. (SPARKS) BD BACTEC¿ FOS¿ (FASTIDIOUS ORGANISM SUPPLEMENT) KIT; SYSTEM, BLOOD CULTURING Back to Search Results
Catalog Number 442153
Medical Device Problem Code Microbial Contamination of Device (2303)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported while using the bd bactec¿ fos¿ (fastidious organism supplement) kit biological contamination was identified.There was no patient involvement or consequences reported.
 
Additional Manufacturer Narrative
Investigation summary: bactec fos supplement kit batch 3249737 is composed of fos supplement batch 3214672 and fos reconstitution fluid batch 3214676.Batch history record review for bactec fos supplement kit batch 3249737 was satisfactory.No notifications were generated during manufacturing and no other complaints have been taken on this batch.Bactec fos supplement is manufactured by rehydrating components in usp purified water and mixed into a homogenous solution.The solution is then sterile filtered, dispensed into vials and stoppered by machine.The vials are lyophilized, and crimp caps are applied per standard operating procedures (sop).Bactec fos reconstitution fluid is manufactured by rehydrating the media components with usp purified water and mixed until a homogenous solution is obtained.The solution is dispensed into vials and capped and crimped per sop.Vials are autoclaved in an air over pressure (aop) autoclave; per manufacturing instructions, using a validated cycle.Post autoclaving, vials are labeled and packaged in a separate packing area.Four fos supplement vials are then manually packaged with one fos reconstitution fluid vial to make a bactec fos supplement kit (material 442153).Batch history record reviews for fos supplement batch 3214672 and fos reconstitution fluid batch 3214676 were satisfactory and no notifications were generated during manufacturing.Qc inspection and testing were satisfactory at time of release.No other complaints have been taken either batch.Retentions of batch 3214676 (reconstitution fluid) were not available; however, batch 3214672 (fos supplement) were available and used for further investigation.There were no observed defects in the available retentions.Two fos supplement vials were reconstituted with sterile water and incubated for 7 days.One was placed into 20-25 degree c incubation, and one was placed into 33-37 degree c incubation.There was no growth observed after incubation.There were no photos or returns submitted to assist with this investigation.This complaint cannot be confirmed.No complaint trends for this defect have been identified for this product; no actions are indicated at this time.Bd will continue to trend for defects.
 
Event or Problem Description
It was reported while using the bd bactec¿ fos¿ (fastidious organism supplement) kit biological contamination was identified.There was no patient involvement or consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ FOS¿ (FASTIDIOUS ORGANISM SUPPLEMENT) KIT
Common Device Name
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key19115473
Report Number1119779-2024-00293
Device Sequence Number12538225
Product Code MDB
UDI-Device Identifier30382904421534
UDI-Public(01)30382904421534
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K920933
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/30/2024
Device Catalogue Number442153
Device Lot Number3249737
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2024
Supplement Date Received by Manufacturer11/04/2024
Initial Report FDA Received Date04/16/2024
Supplement Report FDA Received Date11/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-