| Catalog Number |
9-ASD-019 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Codes |
Cardiac Arrest (1762); Obstruction/Occlusion (2422)
|
| Date of Event |
02/21/2024
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a 19mm amplatzer septal occluder was chosen for implant using an 8f amplatzer trevisio intravascular delivery system.The patient was in sinus rhythm during procedure.The defect measured as 1.6mm on echocardiography.Echocardiographic and fluoroscopic guidance were used during procedure.There was difficulty implanting the device.The device was repositioned several times, and the edge of the inferior vena cava was left out.Finally the edges inside the device were visualized.A stability check was performed, and the device was implanted.Approximately 12 hours after implantation, the device embolized to the patient's left atrium.On 2024, the patient was taken to the procedure room to remove the device, and attempts were made to remove the device from the patient's heart via transcatheter snare.Venous access was made, and a temporary pacemaker was placed and resuscitated.The device had interaction with the mitral valve and travelled to the left ventricular outlet.The device caused an obstruction to blood flow at the left ventricular outlet.The device was attempted to be extracted but was not able to be recovered.The patient entered sustained low output and passed away.The cause of the embolization was believed to be that the edges of the cardiac structures had little firmness.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2024, a 19mm amplatzer septal occluder was chosen for implant using an 8f amplatzer trevisio intravascular delivery system.The patient was in sinus rhythm during procedure.The defect measured as 1.6mm on echocardiography.Echocardiographic and fluoroscopic guidance were used during procedure.There was difficulty implanting the device.The device was repositioned several times, and the edge of the inferior vena cava was left out.Finally the edges inside the device were visualized.A stability check was performed, and the device was implanted.Approximately 12 hours after implantation, the device embolized to the patient's left atrium.On (b)(6) 2024, the patient was taken to the procedure room to remove the device, and attempts were made to remove the device from the patient's heart via transcatheter snare.Venous access was made, and a temporary pacemaker was placed and resuscitated.The device had interaction with the mitral valve and travelled to the left ventricular outlet.The device caused an obstruction to blood flow at the left ventricular outlet.The device was attempted to be extracted but was not able to be recovered.The patient entered sustained low output and passed away.The cause of the embolization was believed to be that the edges of the cardiac structures had little firmness.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of a device embolization into the left atrium 12 hours post implantation and death was reported.Information from the field indicated that there was difficulty implanting the device.A returned device inspection, to rule out any device-related causes, could not be performed as the device was not returned for analysis.However, a review of imaging received from the field and medical review were performed.The reviews revealed the cause of the reported device embolization was likely due to anatomical factors due to difficulty achieving a stable position along the ivc rim; and the cause of the reported death is due to technical difficulty in retrieving the device from the left heart.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Search Alerts/Recalls
|