• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ASCENSION ORTHOPEDICS, INC. RSS GLENOID BASEPLATE IMPACTOR-S; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ASCENSION ORTHOPEDICS, INC. RSS GLENOID BASEPLATE IMPACTOR-S; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number IMP-0961-064-S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/27/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during surgery, the rss glenoid baseplate impactor-s broke.It is unknown how was the procedure completed and if a delay was presented.No injury to patient was reported.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Additional Manufacturer Narrative
H6: the associated device was returned and evaluated.The visual inspection revealed that the plastic tip was broken off of the impactor portion of the device.The black radel material of the impactor had some visible wear and impressions.The visual appearance of the remainder of the device exhibited standard wear with minor surface blemishes.There was corrosion noted on the rod component.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSS GLENOID BASEPLATE IMPACTOR-S
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer (Section G)
ASCENSION ORTHOPEDICS, INC.
11101 metric blvd
austin TX 78758
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19116412
Report Number3002788818-2024-00057
Device Sequence Number7576026
Product Code LXH
UDI-Device Identifier00885556863947
UDI-Public00885556863947
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberIMP-0961-064-S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2024
Supplement Date Received by Manufacturer05/15/2024
10/09/2024
Initial Report FDA Received Date04/16/2024
Supplement Report FDA Received Date05/16/2024
10/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-