| Model Number |
Ø7X 20MM PEEK IF SCRW NON-VENTED |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Foreign Body In Patient (2687)
|
| Date of Event |
02/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 3/21/2024, an fda medwatch notification was received via email.A facility nurse reported that an ar-4020p-07 fastthread peek interference screw was not an adequate fit.As the surgeon removed the screwdriver, it was noticed the peek screw was not at the end of the screwdriver.The surgical site as well as the surrounding areas were searched for the screw but could not be located.An x-ray was taken using fluoroscopy and the screw could still not be located.An mri was taken post-operative which revealed a plastic screw in the patient's patellar pouch.This occurred during a patellar ligament repair procedure on (b)(6) 2024.Additional information received on 3/25/2024: on (b)(6) 2024, the patient underwent revision surgery and had the retained foreign object removed.No known harm to the patient during or after these procedures were completed.No additional information has been provide.Additional information received on 3/26/2024: the revision surgery took place on (b)(6) 2024, and the fastthread peek interference screw was successfully removed from inside the patient.No other arthrex products were explanted and no new products were implanted.The procedure was performed to only remove the fastthread peek interference screw.
|
| |
|
Additional Manufacturer Narrative
|
|
Mw5152261.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause can be attributed to misuse due to misaligned insertion, or prying/leveraging the screw during insertion.
|
| |
|