• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD ALARIS SMARTSITE EXTENSION SET; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD ALARIS SMARTSITE EXTENSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 20127E
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that the set occluded.One sample model 20127e was returned for investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.The set was attached to a bd primary set and primed with 85% glycerin / 15% water.An infusion run was started at a rate of one ml/hr for 24 hrs.There was no observation of fluid blockage.The customer complaint was unable to be replicated.See the attached ¿filtering out the facts: recommendations to optimize performance of in-line filters for parenteral nutrition and intravenous fat emulsion infusions (ivfe)¿ for additional information regarding the use of ivfe with in-line filter administration sets.Although the issue was unable to be replicated and the root cause is unknown bd is looking at opportunities to improve the filter function to alleviate this failure in the future.A device history record review for model 20127e lot number 23129307 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 20dec2023.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.
 
Event or Problem Description
No additional info.
 
Event or Problem Description
It was reported that bd alaris smartsite extension set was occluded the following information was received by the initial reporter with the following verbatim: lipids (in bag) kept alarming occluded.
 
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE EXTENSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key19116964
Report Number9616066-2024-00573
Device Sequence Number7141032
Product Code FPA
UDI-Device Identifier07613203019248
UDI-Public(01)07613203019248
Combination Product (Y/N)N
Initial Reporter StateAK
Initial Reporter CountryUS
PMA/510(K) Number
K022209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number20127E
Device Lot Number23129307
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2024
Supplement Date Received by Manufacturer05/10/2024
Initial Report FDA Received Date04/16/2024
Supplement Report FDA Received Date05/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/20/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-