| Catalog Number |
05P58-25 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
Anxiety (2328)
|
| Date of Event |
04/08/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2024, abbott point of care (apoc) was contacted by a customer regarding i-stat b-hcg cartridges that yielded positive results on a 10 year old female patient.There was no additional patient information at the time of this report.Return product is available for investigation.Method date collected tested result (iu/l) sample i-stat (b)(6) 2024 08:20 immediately 39.8 wb.Cobas (b)(6) 2024 ni 09:37 <1.0 plasma.I-stat (b)(6) 2024 08:20 15:00 51.4 plasma.Cobas (b)(6) 2024 ni 15:30 <1.0 plasma.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.I-stat positive cut-off values: b-hcg < 5.0 iu/l negative.5.0 < ss-hcg < 25.0 iu/l indeterminate.B-hcg >25.0 iu/l positive.Intended use: the i-stat® total beta-human chorionic gonadotropin (b-hcg) test is an in vitro diagnostic test for the quantitative and qualitative determination of b-hcg in venous whole blood or plasma samples using the i-stat 1 analyzer systems.The test is intended to be used as an aid in the early detection of pregnancy and is for prescription use only.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 29-apr-2024.A review of the device history record confirmed the cartridge lot met finished goods release criteria.Cartridge testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.An (product complaint level 2 and level 3 investigation procedure).No deficiency has been identified for total b-hcg cartridge lot f23311a.
|
| |
|
Search Alerts/Recalls
|