| Catalog Number |
04625315160 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The serial number for meter a was (b)(6).The serial number for meter b was (b)(6).The product was requested for investigation.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable results from coaguchek xs plus meter compared to a laboratory result using another coaguchek xs plus meter.The result from meter a was 3.2 inr.The result from meter b was 2.4 inr.The results were obtained less than 4 hours apart.The patient¿s therapeutic range was 2.0-3.0 inr.The patient¿s testing frequency is every 5-6 weeks.
|
| |
|