| Model Number |
VTICM5_13.2 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted an implantable collamer lens into the patient's eye.The surgeon reports the vault is extremely high.A lens exchange was planned.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - the reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2; -12.50/3.0/130 (sphere/cylinder/axis) implantable collamer lens into the patient's right eye (od).The surgeon reports there is excessive vaulting with significant reduction or irido-corneal angles, shallow ac, and the patient also had ifis during surgery.On (b)(6) 2024 the lens was explanted and this resolved the problem.The cause of the event was reported as the device.H6 - health effect clinical code: 4581 - significant reduction of irido corneal angles, ifis, shallow ac.H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim#: (b)(4).
|
| |
|