• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_13.2
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
Health Effect - Clinical Codes Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
The reporter indicated that the surgeon implanted an implantable collamer lens into the patient's eye.The surgeon reports the vault is extremely high.A lens exchange was planned.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
B5 - the reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2; -12.50/3.0/130 (sphere/cylinder/axis) implantable collamer lens into the patient's right eye (od).The surgeon reports there is excessive vaulting with significant reduction or irido-corneal angles, shallow ac, and the patient also had ifis during surgery.On (b)(6) 2024 the lens was explanted and this resolved the problem.The cause of the event was reported as the device.H6 - health effect clinical code: 4581 - significant reduction of irido corneal angles, ifis, shallow ac.H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key19119103
Report Number2023826-2024-01806
Device Sequence Number15516041
Product Code MTA
UDI-Device Identifier00840311312351
UDI-Public00840311312351
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVTICM5_13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2024
Supplement Date Received by Manufacturer04/28/2024
Initial Report FDA Received Date04/16/2024
Supplement Report FDA Received Date04/30/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL-SFC-45, LOT#-UNK; FOAM TIP PLUNGER MODEL-FTP, LOT#-UNK; INJECTOR MODEL-MSI-PF, LOT#-UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-