| Catalog Number |
MX960 |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Failure to Zero (1683); Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1 - phone number: (b)(6).H3 - other: device has not been returned to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the "pressure plate zero is not within 100±3.The p1 p2 p3 p4 value shows 196-197.The port is not in good contact and cannot be connected.The pressure value is displayed as 1067-1066; which is marked as 100±3." there was no patient involvement reported and no harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction-d9 - date returned to mfg-5/2/2024 one device was returned for evaluation.Visual inspection does not show any signs of physical damage.Functional testing failed to meet specifications for multiple test specification criteria.Additional evaluation completed by dis-assembly of the two halves of the transducer housings to visually inspect internal assembly of the transducer, it was found to have incomplete solder application on one of the input impedance connections to the pc board allowing intermittent connection.Problem for port is not in good contact and cannot be connected was unable to be replicated.The customer's problem was confirmed.The root cause was the manufacturing assembly process for solder application.No further actions.The device history review of the reported lot number found no non-conformities during the manufacturing process.
|
| |
|