• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA ARTICULAR SURFACE FXD BRG (CPS) RIGHT 12 MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA ARTICULAR SURFACE FXD BRG (CPS) RIGHT 12 MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Pain (1994)
Date of Event 02/06/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Additional associated products: 42504607002 persona femur cemented std right size 11 lot# 64532701.42545001012 persona femoral central cone size medium lot# 64189936.42556807005 persona femoral post augment cemented size 11, 11+5mm lot# 64485584.42556607005 persona femoral dist augment cemented size 11, 11+5mm lot# 64664231.42560113520 persona stem ext straight splined uncemented 20mm dia 135mm l lot# 64635804.42542007902 persona tibia fixed cemented right size g lot# 64334603.42560613514 persona stem ext 6mm offset splined uncemented 14mm dia 135mm l lot# 64729231.42545000513 persona tibial central cone size large lot# 64189641.
 
Event or Problem Description
It was reported a morbidly obese patient began to experience an increase in pain and reports taking a 'couple tumbles' approximately 8 months post implantation.Patient presented to the er at this time.During this visit the patient reports running out of norco as it was being taken more than prescribed for the pain.A ct scan was performed and displayed no abnormal findings and the patient was sent home with over-the-counter nsaids and topical analgesics.All implants remain in place and the patient has continued within the study without further complication to the study knee.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, g3, g6, h1, h2, h3, h6, h10, h11.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A ct scan was performed and displayed no abnormal findings.No further analysis can be made.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The reported products were reviewed for compatibility with no issues noted.An adequate complaint history review cannot be performed as the complication, failure mode, and/or harm experienced by the user is unknown.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: the event is for pain and multiple falls with unknown cause and is considered a serious injury as the illness or impairment required hospitalisation, medical intervention and/or surgical intervention.Discharge instructions: nsaids, otc topical analgesics, heat/ice.This complaint cannot be confirmed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA ARTICULAR SURFACE FXD BRG (CPS) RIGHT 12 MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19119286
Report Number3007963827-2024-00119
Device Sequence Number12819431
Product Code JWH
UDI-Device Identifier00889024244795
UDI-Public(01)00889024244795(17)241231(10)64528882
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K123459
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model NumberN/A
Device Catalogue Number42522601012
Device Lot Number64528882
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2024
Supplement Date Received by Manufacturer09/10/2024
Initial Report FDA Received Date04/16/2024
Supplement Report FDA Received Date09/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARATIVE.
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient SexMale
Patient Weight171 KG
-
-