• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. FEMORAL STEM 12/14 NECK TAPER; PROSTHESIS, HIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. FEMORAL STEM 12/14 NECK TAPER; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870)
Date of Event 03/27/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D1: femoral stem 12/14 neck taper extra ext.Offset size 2 130 mm stem length packaged with standard centralizer.G2: foreign: australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient suffered a periprosthetic fracture from a fall five (5) weeks post operative.The surgical technique of the product was utilized, and the implant did not contribute to the fall.The stem remains stable and was not revised.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.
 
Event or Problem Description
No additional event information to report at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.Proposed component code: mechanical (g04)- stem.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: oblique displaced fracture of the left femur near the tip of the femoral implant as noted it was reported the patient suffered a periprosthetic fracture due to a fall.As the reason for the fall is unknown, a definitive root cause cannot be determined.This complaint was confirmed based on the provided x-rays confirming the reported event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL STEM 12/14 NECK TAPER
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19119325
Report Number0002648920-2024-00105
Device Sequence Number15524637
Product Code JDI
UDI-Device Identifier00889024145801
UDI-Public(01)00889024145801(17)310908(10)65047403
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K191735
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number00811400230
Device Lot Number65047403
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/27/2024
Supplement Date Received by Manufacturer04/17/2024
06/11/2024
Initial Report FDA Received Date04/16/2024
Supplement Report FDA Received Date05/06/2024
06/13/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient Age74 YR
Patient SexMale
Patient Weight115 KG
-
-