It was reported that outside of surgery, when two brand new a.T.S 3200 units were powered on and tourniquet cuffs attached, the pressure fluttered and could not establish stabilized pressure.They tried several cuffs same result.They did a leak test on cuffs and they did not show a leak.There was no patient involvement.Due diligence is complete, no further information is available.
|
|
This complaint is recorded by zimmer biomet under (b)(4).Following sections were updated or corrected: b4, b5, d2, d4, g1, g3, g6, h1, h2, h3, h4, h6, and h11.Review of the most recent repair record identified no repairs related to the reported event.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|