• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR CUP Back to Search Results
Catalog Number UNK HIP ACETABULAR CUP
Device Problem Osseointegration Problem (3003)
Patient Problems Foreign Body Reaction (1868); Pain (1994); Inadequate Osseointegration (2646)
Event Date 03/15/2024
Event Type  Injury  
Event Description
Voluntary report (mw5153031) received revised total hip arthroplasty due to painful loosening.Discovered evidence of metallosis.Depuy srom stem ceramic cup, polyethylene liner, pinnacle acetabular cup.Failed tha only 3.5 years old due to metallosis.
 
Manufacturer Narrative
Product complaint (b)(4).D4: the device catalog number is unknown; therefore, udi is unavailable.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
An email was received from the patient.It stated that the implant caused metallosis, proven by a hip revision surgery.Brown/black fluid pulled from joint space.This caused stem loosening.The implant didn¿t even make it to it's 4th year.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP ACETABULAR CUP
Type of Device
HIP ACETABULAR CUP
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key19120909
MDR Text Key340338491
Report Number1818910-2024-08448
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP ACETABULAR CUP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/09/2024
Initial Date FDA Received04/16/2024
Supplement Dates Manufacturer Received04/25/2024
Supplement Dates FDA Received04/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNK HIP ACETABULAR LINER METAL; UNK HIP FEMORAL HEAD METAL; UNK HIP FEMORAL SLEEVE SROM; UNK HIP FEMORAL STEM SROM
Patient Outcome(s) Required Intervention;
-
-