| Model Number |
DAC-NT |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
E1: reporter phone number: (b)(6).Unique device identifier (udi #): the udi is unknown because the lot number was not provided.During processing of this incident, attempts were made to obtain complete device information.Further information was requested but not received.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 3005334138-2024-00142 it was reported that during a procedure a message was received stating the electrode has been changed.As such, the treatment was stopped, and the patient was changed to medication treatment.Reportedly, there were no adverse consequence to the patient.A replacement adapter and same needle were used but the issue persisted.However, when troubleshooting with a representative¿s adapter cable, the issue was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|