| Catalog Number |
EIC8872-01 |
| Medical Device Problem Codes |
Break (1069); Material Erosion (1214)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an ent procedure, after opening the precise wand package and using it for 5 minutes, the electrode wire broke.The broken pieces were removed from the patient using suction.The procedure was successfully completed using a back-up device.There was a surgical delay less than 30 minutes and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device was received for evaluation.A visual inspection revealed no manufacturing abnormalities.Fluid is in the fluid line.Electrodes have been heavily used.All of the electrodes have eroded.The wand is bent from use.Bio debris is present.Product was out of the original packaging.No packaging returned.A functional evaluation was performed on the returned device and found that it registered settings (7,3).Coagulation and plasma were generated as intended with the available electrode pieces.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that photos of the device were provided for review and confirm the reported.The patient was noted to be a 5-year-old female.No patient injuries or adverse consequences were reported.The patient¿s health status was reported as healthy.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include hitting the device tip against a hard surface such as an insertion cannula, using the device as a lever to enlarge surgical site, mechanical displacement of tissue through applied force or activating the device above recommended settings for prolonged periods of time.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|