• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PERFUSION SYSTEMS EOPA ARTERIAL CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS EOPA ARTERIAL CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO Back to Search Results
Model Number 77424
Medical Device Problem Codes Fluid/Blood Leak (1250); Mechanical Problem (1384)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 03/25/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information that during use of an eopa arterial cannula, it was reported that there was a small amount of patient blood loss from the device during cannulation.There was approximately 100-200mls of blood loss.The customer stated that the white stylus appeared too small/narrow for the cannula which resulted in a messy cannulation. the customer stated that they are an experienced cannulator.There was no transfusion required.The cannula was not heated or cooled prior to cannulation.The patient was already heparinised, therefore, the blood was returned to the extracorporeal bypass circuit and then to the patient circulation once the bypass circuit was connected to the aortic cannula.The device was used.There was no adverse patient effect associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EOPA ARTERIAL CANNULA
Common Device Name
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key19123139
Report Number2184009-2024-00188
Device Sequence Number20738377
Product Code DWF
UDI-Device Identifier00763000135645
UDI-Public00763000135645
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K031037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/19/2026
Device Model Number77424
Device Catalogue Number77424
Device Lot Number2023070728
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2024
Initial Report FDA Received Date04/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-