| Catalog Number |
0284514 |
| Medical Device Problem Codes |
Break (1069); Deflation Problem (1149); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a valvuloplasty procedure, the ptv balloon allegedly partially separated at the shaft and the balloon allegedly wasn't completely deflated.It was further reported that the actual balloon was still connected via the lumen to fill the balloon.Furthermore, it was allegedly hard to remove through the sheath.Reportedly, it was removed all together and the procedure was completed using another balloon.There was no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiration date: 07/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
| |
|
Event or Problem Description
|
|
It was reported that during a valvuloplasty procedure, the ptv balloon allegedly partially separated at the shaft.It was further reported that the actual balloon was still connected via the lumen to fill the balloon.Furthermore, it was allegedly hard to remove through the sheath.Reportedly, it was removed all together and the procedure was completed using another balloon.There was no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one true ptv dilatation catheter was received for evaluation.A complete circumferential break to the distal end of the catheter shaft proximal to balloon was noted.During visual analysis, the balloon was noted to be separated from the catheter shaft at the proximal end leaving the inner guidewire lumen exposed.No other anomalies were noted.Due to the condition of the device, no functional testing was performed.Based on the received sample visual analysis, a catheter shaft break with the inner guidewire lumen in the proximal end of balloon could be observed, further the balloon was attached with the inner guidewire lumen moved distal to the balloon could be noted.Therefore, the investigation is confirmed for the reported break and difficult to removal issues.A definitive root cause for the reported break and difficult to removal issues could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 07/2026).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|