| Model Number |
DDMC3D4 |
| Medical Device Problem Codes |
Failure to Deliver Shock/Stimulation (1133); Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Abdominal Distention (2601); Heart Failure/Congestive Heart Failure (4446); Peripheral Edema (4578)
|
| Date of Event |
03/07/2024
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead triggered lead integrity alert (lia) for sensing integrity counter (sic) and high rate non-sustained episodes.The rv lead was also stated to be oversensing.It was further reported that the lia was suspected to be triggered due to t-wave oversensing (twos), and an atrial high rate episode was labeled ventricular tachycardia (vt) due to oversensing of noise.It was further reported that the patient's implantable cardioverter defibrillator (icd) inappropriately labeled an episode of sustained vt as supra ventricular tachycardia (svt).The patient had reported shortness of breath, abdominal distention and peripheral edema.After a family consultation, the device was deactivated with use of a magnet.The patient passed away the next day.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory had an observation relating to supra ventricular tachycardia (svt) detection.Analysis of the device memory indicated the device failed to deliver a shock.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|