| Model Number |
21231-6 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
03/25/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
We have inspected the dhf of the lot and the material certificate and the batch was processed through the normal manufacturing and inspection operations with no non-conformances or rework noted.The article met all dimensional and visual criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Based on the information available, no final conclusion could be drawn and it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.This report does not reflect a conclusion by fda, i.T.S.Gmbh or its employees that the report constitutes an admission that the device caused or contributed to the potential event described in this report.This is report 1/2.
|
| |
|
Event or Problem Description
|
|
It was reported that a non-locking screw backed out approx.5 weeks after surgery.In addition the two distal locking screws show signs of loosening in the bone, but are still locked in the plate.A second surgery was necessary.
|
| |
|
Search Alerts/Recalls
|