| Model Number |
MAJ-2315 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Full e1 establishment name: social medical corporation (b)(6) hospital.During device evaluation at olympus, it was found there was a crack at the rear end of the cover.A review of the device history record found no deviations that could have caused or contributed to the reported issue.A definitive root cause of the broken cover could not be determined; however, it's possible that increased stress applied to the distal cover due to semi-insertion or release of stress applied in the installed state due to some chemicals caused cracking over time in areas of weak strength or increased stress.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the instructions for use (ifu).If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
|
| |
|
Event or Problem Description
|
|
The customer reported the underside of the opening on the single use distal cover was torn.The issue occurred when covering the endoscope during preparation for use.Another cover was used to complete the procedure.There was no reported patient harm or impact due to this event.
|
| |
|