| Model Number |
N/A |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a knee procedure, the surgeon noted debris in the sterile packaging of the stem prosthesis.Another stem was used to complete the procedure.No adverse events have been reported as a result of the malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).G2 foreign source: finland customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event could not be confirmed.Visual examination of the returned product identified an unknown debris.The unknown debris was sent for analytical testing, which determined the debris is consistent with skin.As the packaging was previously opened, the source of the skin cannot be confirmed.Device history record was reviewed and no discrepancies were found.Review of complaint history found no additional related issues for the reported part and part lot combination.A definitive root cause cannot be determined.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|