| Model Number |
MAJ-2315 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported the single use distal cover was opened and found to be defective.The issue occurred during a therapeutic endoscopic retrograde cholangiopancreatography procedure.Another cover was used to complete the procedure.There was no reported patient harm or impact due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The dhr was unable to be reviewed for this device since the lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.A definitive root cause of the broken cover could not be determined; however, it's possible that increased stress applied to the distal cover due to semi-insertion or release of stress applied in the installed state due to some chemicals caused cracking over time in areas of weak strength or increased stress.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the instructions for use (ifu).If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
|
| |
|