| Catalog Number |
VASC1446 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
According to the customer, "it really does seem to be a mixture of pre-procedure and middle of the procedure occurrences of the same issue happening where the syringe will not stay on the manifold without physically being held in place".There is no further information.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.In an abundance of caution, and in response to an fda 483 issued for cfn 1417592 on 22-jan-2024, this medwatch is being filed for the reported problem/issue.If additional relevant information becomes available a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
According to the customer, "it really does seem to be a mixture of pre-procedure and middle of the procedure occurrences of the same issue happening where the syringe will not stay on the manifold without physically being held in place".
|
| |
|
Additional Manufacturer Narrative
|
|
Update to b3, d9, h3, and h6: after further investigation, it was determined that the date of the event was 01/16/2024.A sample was returned to and evaluated by the manufacturer on 03/08/2024.The investigation included trend analysis, testing of the actual/suspected device, and a review of production records.The investigation identified a mechanical issue in which the syringes backed off/rolled free of the manifold connection, thereby replicating the reported defect.A cause remains unestablished.If any additional relevant information is identified or obtained, a supplemental medwatch will be submitted.
|
| |
|
Search Alerts/Recalls
|