| Model Number |
SUTR ANCH,BIO-COMP P-LCK,2.9X 12.5MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 3/22/2024, it was reported by a sales representative via email that an ar-2923bc biocomposite pushlock anchor split in half during insertion.The broken fragments were removed, and another anchor was used to complete the case.This was discovered during a slap repair procedure on 3/21/2024.Additional information received on 3/27/2024: the case was not delayed and additional anesthesia was not needed.The patient suffered no negative effects.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint allegation is confirmed.One unpackaged ar-2923bc serial/batch number (b)(6) was received for investigation.The visual evaluation showed that the anchor had broken off, with only a small fragment returned for evaluation.Functional testing was not performed due to the damage to the received device.The method of bone quality encountered, was described as a hard.The most likely cause(s) for the reported failure can be user error, improper bone preparation, misaligned insertion, or excessive force by leveraging/prying the device.According to dfu-0087 at revision 13.E.Warnings.6.Bio-absorbable only: attempting implantation into hard, dense bone and/or drilling/punching smaller diameter holes than recommended may cause failure (breakage) of the implant during insertion.G.Precautions.4.Pushlock and swivelock suture anchors only: during anchor insertion, ensure that the angle of anchor insertion is coaxial to that of the previously prepared bone socket.
|
| |
|
Search Alerts/Recalls
|