| Catalog Number |
1012625-29 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat the renal artery.The 9.0x29mm omnilink elite balloon expandable stent was attempted to be stented within another unspecified stent; however, the stent dislodged and had to be snared.Clinically significant delay was noted as the stent had to be snared.There was no adverse patient sequela.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no other complaints.The investigation determined the reported difficulties and subsequent unexpected medical intervention appear to be related to circumstances of the procedure, as it is likely the device interacted with the previously implanted stent during advancement, causing the reported failure to advance.Further interaction with the previously implanted stent during retraction of the device likely contributed to reported difficult to remove, ultimately causing the reported stent dislodgement.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
Subsequently, after the initial reported was filed it was noted that the stent became stuck with the previously implanted and dislodged.The stent had to be snared.Another device was used to successfully complete the procedure.No additional information was provided.
|
| |
|