ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1012625-29
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat the renal artery.The 9.0x29mm omnilink elite balloon expandable stent was attempted to be stented within another unspecified stent; however, the stent dislodged and had to be snared.Clinically significant delay was noted as the stent had to be snared.There was no adverse patient sequela.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.Production record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no other complaints.The investigation determined the reported difficulties and subsequent unexpected medical intervention appear to be related to circumstances of the procedure, as it is likely the device interacted with the previously implanted stent during advancement, causing the reported failure to advance.Further interaction with the previously implanted stent during retraction of the device likely contributed to reported difficult to remove, ultimately causing the reported stent dislodgement.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Event or Problem Description
Subsequently, after the initial reported was filed it was noted that the stent became stuck with the previously implanted and dislodged.The stent had to be snared.Another device was used to successfully complete the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key19126872
Report Number2024168-2024-04781
Device Sequence Number18049600
Product Code NIO
UDI-Device Identifier08717648179334
UDI-Public(01)08717648179334(17)250731(10)2022241
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Catalogue Number1012625-29
Device Lot Number2022241
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/26/2024
Supplement Date Received by Manufacturer05/20/2024
Initial Report FDA Received Date04/17/2024
Supplement Report FDA Received Date06/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-