|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3542251 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
02/21/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast prosthesis rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast augmentation procedure with mentor memorygel breast implant 225cc gel breast prostheses that both ruptured post implantation.As a result, the patient underwent bilateral explantation on an unknown date.This medwatch report is for the patient¿s right breast prosthesis.The reported implantation date is after the explantation date of the suspect medical device.Mentor will report the implantation date as received.If clarification is received, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
On 08-may-2024, mentor completed the investigation on the suspect medical device.The investigation was completed on a photo of the suspect medical device because the physical device has not been received by mentor.Device evaluation summary: according to the information received, the patient experienced a rupture in the breast implant.Upon visual evaluation of the image provided in the complaint, gel exposure was observed on the implant.As the product involved in this complaint was not received, the device couldn¿t be analyzed according to our procedures.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: in the initial report, block b5 stated ¿the reported implantation date is after the explantation date of the suspect medical device.¿ it should read that the implantation date is after the expiration date of the suspect medical device.On 27-may-2024, mentor received additional information about the event.The explantation date is (b)(6) 2024.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 27-jun-2024, the mentor failure analysis lab received the device for evaluation.Additionally, mentor became aware of the following: - the patient dob is (b)(6) 1969.- the patient underwent a breast augmentation revision procedure with the suspect medical device.- the patient¿s explanted breast prostheses were replaced with mentor memorygel xtra breast implant 240cc gel breast prostheses.On 28-jun-2024, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced a rupture in the breast implant.The product was returned to mentor for evaluation.Mentor conducted visual inspection of the returned device.Visual analysis of the returned sample revealed that the breast implant was found to have an area of delamination at the union between the patch and shell, causing gel leakage.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Although potential causes of patch delamination are unknown, stresses and forces on the implant in-vivo or exposure to betadine at insertion could result in delamination and ruptures.Per the current product insert data sheet (pids), rupture is a known complication associated with these devices and can occur at any time during or after implantation.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|