The following devices were also listed in this report: 180506 mck femoral-lm-rl-sz 6; lot 2gv1-1 180606 mck tibial baseplate-lm/rl-sz 6; lot 26360821-01 it cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Reported event: an event regarding pain involving a mako insert was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|