• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARL ZEISS MEDITEC PRODUCTION LLC CT LUCIA; CT LUCIA 621P

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARL ZEISS MEDITEC PRODUCTION LLC CT LUCIA; CT LUCIA 621P Back to Search Results
Model Number 621P
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 03/20/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been returned.Thus, a proper device analysis could not be completed nor the reported issue confirmed.At this time, without a proper device analysis, a definitive root cause cannot be determined.However, there is no indication of a product quality issue with respect to manufacturing, design, or labeling of the lens.Based on the information provided, the risk assessment for the lucia product, and our experience we have determined that the following factors may have cause or contributed to the reported issue is but is not limited to: -misplacement of the lens -inadequate application of ovd -dwell time after preparation the device history records were reviewed and there were no non-conformities or deviations noted during the manufacturing of the lens that would have caused or contributed to the nature of this complaint.Our lenses are 100% inspected before they leave our manufacturing site.Therefore, we are confident that the lens was processed per standard operating procedures and inspections and met all criteria for release.
 
Event or Problem Description
It was reported that the lens was not deploying correctly, which was in the opposite direction than normal.The lens was injected into the eye partially and then had to be removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CT LUCIA
Common Device Name
CT LUCIA 621P
Manufacturer (Section D)
CARL ZEISS MEDITEC PRODUCTION LLC
1040 s. vintage ave. bld. a
ontario CA 91761
Manufacturer (Section G)
CARL ZEISS MEDITEC PRODUCTION LLC
1040 s. vintage ave. bld. a
ontario CA 91761
Manufacturer Contact
manjaya hegde
5300 central parkway
dublin, CA 94568
9252164697
MDR Report Key19129854
Report Number3010126268-2024-00047
Device Sequence Number20558086
Product Code HQL
UDI-Device Identifier00843045102373
UDI-Public(01)00843045102373(11)230215(17)260131
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P10016-S008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model Number621P
Device Catalogue Number003500-0056-491
Device Lot Number3S230173
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2024
Initial Report FDA Received Date04/17/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-