UNITED ORTHOPEDIC TIBIAL TOWER, NON-CEM, #3-#5; TIBIAL GUIDE Back to Search Results
Model Number 9404-2902
Medical Device Problem Codes Break (1069); Material Separation (1562); No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The problem may be caused by an inappropriate usage.While encountering to dense bone, the surgeon should use the tibial peg drill to make four pilot holes before positioning the tibial tower onto the tibial baseplate trial, otherwise the device breakage might happened.On the other hands, the device manufacturing manner of the spike part has been revised from one-piece machining to welding to further enhance the device strength.
 
Event or Problem Description
Dr.Removed the tibial tower using the slotted hammer with the punch and punch adapter inside of the tower.After handing off the assembly to the surgical tech, the tech noticed one of the spikes had broken off and was lying beside the assembly on the mayo stand.Standard surgical protocol was followed, there was no delay to the case, the patient was not affected and the case was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL TOWER, NON-CEM, #3-#5
Common Device Name
TIBIAL GUIDE
Manufacturer (Section D)
UNITED ORTHOPEDIC
no. 57, park ave. 2
science park
hsinchu city, 30075
TW  30075
Manufacturer (Section G)
UNITED ORTHOPEDIC
no. 57, park ave. 2
science park
hsinchu city, 30075
TW   30075
Manufacturer Contact
lois ho
no. 57, park ave. 2
science park
hsinchu city, 30075
TW   30075
MDR Report Key19131655
Report Number9681642-2024-00014
Device Sequence Number17867690
Product Code LXH
UDI-Device Identifier04719886892123
UDI-Public(01)04719886892123(10)22G859
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9404-2902
Device Lot Number22G859
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2024
Initial Report FDA Received Date04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-