| Model Number |
10-80-00 |
| Medical Device Problem Code |
Pumping Stopped (1503)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Livanova deutschland received a report that , during installation, the s5 single roller pump 150 pump stopped and displayed alarm with yellow triangle shape including exclamation mark.No other activity was possible when the alarm was displayed.At the time of the first occurrence of the problem, the message 'motor controller failed pump 4' was displayed and, afterward, the message was not longer displayed.Additional tests found the problem recurs intermittently regardless of the operating state.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 roller pump.The incident occurred in south korea.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
Through follow up, livanova was informed the problem with the roller pump was a problem during the installation process and the pump was not in use.As a result, the pump was not installed in the hospital and the pump was returned to the distributor.The event has been re-assessed as not reportable since issue was related to an out of box failure and would have been detected anyway prior device use on a patient through the functional tests to be performed at installation phase.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
| |
|
Event or Problem Description
|
|
See initial report.
|
| |
|