• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC CAVITRON SELECT RESERVOIR(DNA) G123; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC CAVITRON SELECT RESERVOIR(DNA) G123; SCALER, ULTRASONIC Back to Search Results
Catalog Number G123
Medical Device Problem Codes Restricted Flow rate (1248); No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
While using a cavitron select with reservoir g123 they allege that they have no water and the handpiece gets hot, no injury resulted.
 
Additional Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Additional Manufacturer Narrative
"04/18/24 evaluation tech: tl emn hp; cable, conn/gun cable kink/pinch/twist g124 -the poor water flow due to twisted handpiece cable causing handpiece gets hot.Worn steri-mate handpiece.Recommend checking all inserts being used are not worn, corroded, damaged and /or clogged and handpiece cable's lavage control is properly adjusted also, follow instructions to bleed the air from the handpiece when installs insert.Make sure screw the cap clockwise to create an airtight seal when using reservoir pump.G123 - no issues found.Unit operates properly.Will replace damaged/worn components and recalibrate unit to factory specs upon estimate approval.Handpiece # 202110 handpiece cable # 11210.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON SELECT RESERVOIR(DNA) G123
Common Device Name
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
dan eagar
221 w. philadelphia st.
york, PA 17401
7178494593
MDR Report Key19132624
Report Number2424472-2024-00030
Device Sequence Number6891334
Product Code ELC
UDI-Device IdentifierD003813081
UDI-PublicD003813081
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K970123
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 05/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberG123
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/16/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/18/2024
Supplement Report FDA Received Date05/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-