| Model Number |
21987 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Corneal Edema (1791)
|
| Date of Event |
03/20/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Although requested, the device was not returned for evaluation and additional information regarding the event was not provided.A review of the device history record (dhr) did not identify any anomalies or nonconformities that could be related to this event.The lot history, trend analysis, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, the root cause of this event could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that during implantation of an intraocular lens (iol), after inserting the iol, the surgeon noticed the iol optic was damaged.The incision was enlarged and an intraoperative lens exchange was performed using a backup iol.Corneal edema was reported.Additional information was requested, but not received.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation.Microscopic examination found the shuttle attached correctly to the device.No damage was observed.
|
| |
|
Event or Problem Description
|
|
Additional information was received indicating that the incision size was enlarged to 2.5mm, which is within standard range for incision enlargement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on additional information received this event no longer meets reportability requirements and is no longer deemed reportable.
|
| |
|