| Catalog Number |
118000 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial left shoulder arthroplasty approximately six (6) weeks ago.Subsequently, the patient was seen by their surgeon three (3) weeks later and it was discovered that the implants had disassociated.The patient underwent a revision surgery approximately three weeks ago.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: medical products: item#: 110027734, compr vrs glen pps min tpr adr; lot#: 66573916.Item#: 110030776, cr 40mm glenosphere std cocr; lot#: 65398246.H3: product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional and corrected information.The following sections were corrected: a3; d4.Visual examination of the returned product/provided pictures identified one 25mm versa-dial taper adaptor item# 118000 lot#j7614800 was returned and evaluated.Upon visual inspection the taper adapter there are galling to post.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: there is a reverse-type left shoulder arthroplasty.The glenosphere is disassociated and displaced from the glenoid baseplate and the proximal humerus is anterosuperiorly subluxed/dislocated in relation to the glenoid.Operative change is noted at the acromioclavicular joint and there is regional heterotopic ossification.There is no fracture or evidence of implant loosening.Bone quality is osteopenic.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|