| Catalog Number |
110027734 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
03/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: medical products: item#: 118000, 25mm versa-dial taper adaptor; lot#: j7614800 item#: 110030776, cr 40mm glenosphere std cocr; lot#: 65398246 h3: product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial left shoulder arthroplasty approximately six (6) weeks ago.Subsequently, the patient was seen by their surgeon three (3) weeks later and it was discovered that the implants had disassociated.The patient underwent a revision surgery approximately three weeks ago.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional and corrected information.The following sections were updated.The following sections were corrected: a3; d10 visual examination of the returned product/provided pictures identified one compr vrs glen pps min tpr adr item# 110027734 lot# 66573916 was returned and evaluated.Upon visual inspection the vrs has scuffing on the face of the device.The female taper of the vrs has galling to the inner taper.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: there is a reverse-type left shoulder arthroplasty.The glenosphere is disassociated and displaced from the glenoid baseplate and the proximal humerus is anterosuperiorly subluxed/dislocated in relation to the glenoid.Operative change is noted at the acromioclavicular joint and there is regional heterotopic ossification.There is no fracture or evidence of implant loosening.Bone quality is osteopenic.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|