| Catalog Number |
366882 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that during use with the bd vacutainer® sst¿ ii advance plus blood collection tubes, an unspecified number of tubes had fibrin present in the sample after centrifugation.There was no report of patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: bd had not received samples, but 6 photos were provided for investigation.The photos were reviewed and the indicated failure mode for fibrin was observed.Additionally, 100 retention samples from bd inventory were evaluated by visual examination and the issue of fibrin was not observed.Complaints for sample quality are under statistical control for the month of (b)(6) 2024.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode fibrin.Bd was not able to identify an exact root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|
Event or Problem Description
|
|
It was reported that during use with the bd vacutainer® sst¿ ii advance plus blood collection tubes, an unspecified number of tubes had fibrin present in the sample after centrifugation.There was no report of patient impact.
|
| |
|