| Model Number |
WA50011A |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device evaluation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the rigid video scope displayed a blurry image.There was no patient involvement.Event description-scope is blurry.Cutover y/n-y.When during procedure did event occur?-unknown ^09.
|
| |
|
Event or Problem Description
|
|
It was reported the rigid video scope displayed a blurry image.There was no patient involvement.Event description-scope is blurry cutover y/n-y when during procedure did event occur?-unknown ^09.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Please see also section h2, h3, h4, h6 and h11 for updates.The device was returned to olympus for evaluation and the customer's allegation was confirmed.In addition, new damages/defects were observed: debris and moisture found under distal cover glass, and dents on distal edge.Based on the results of the investigation, the most probable cause of the complaint is cause traced to component failure, which is expected or random component failure without any design or manufacturing issue a device history record- dhr review was conducted, and no issues were identified.Olympus will continue to monitor field performance for this device.
|
| |
|