It was reported that the sphincterotome videoscope exhibited: during a therapeutic endoscopic retrograde cholangiopancreatography procedure, when opening the sphincterotome and testing it, the knife wire was damaged.There was no error message.The procedure was delayed for 18 minutes, and the patient was already under general anesthesia.The product was replaced with one of the same model and lot to finish the procedure.There were no reports of patient harm, and the patient was stable.No medical intervention was required.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Additionally, to provide a correction to the initial with information inadvertently left out (b5 and g2) and to provide a correction to the initial field (e3).The subject device was manufactured on jan 2023 based on the provided 3 digit lot information.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation and past investigation results, it is likely the mechanism causing the breakage of the cutting wire might be the following: 1) the device was energized in one of these following situations.- the cutting wire is in point contact with tissue or they were being close to each other.- the cutting wire is in point contact with distal end of endoscope or they were being close to each other.2) an electrical discharge occurred in the cutting wire, and the cutting wire became hot instantly.3) the cutting wire was broken.However, the root cause of the reported malfunction could not be identified.The event can be prevented by following the instructions for use which state: "since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result." "be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.In case the cutting wire contacts the forceps elevator, insufficient output or unintended tissue injury may occur." "do not activate output while the cutting wire touches the metal parts of the endoscope, or they are being close together.This could burn the tissue and/or damage the endoscope or the instrument." olympus will continue to monitor field performance for this device.
|