| Model Number |
A22251A |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1.Full establishment name: (b)(6) hospital.The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported, the hf-resection electrode, roller, 24-28 fr., 12° and 30° displayed a bent electrode tip.The issue occurred during preparation for use for a transcervical resection procedure.The procedure was completed using the same set of equipment.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.New information added to the following fields: d8, d9, h3, h4, h6, h11.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over two (2) years since the subject device was manufactured.The device was returned, and an evaluation was completed for it.During inspection, olympus confirmed the reported event.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the defect to the product was caused by improper handling during preparation at the customer¿s site or whether there was an undetected fault from the production process.Olympus will continue to monitor field performance for this device.
|
| |
|