• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA25-100/I16-40
Medical Device Problem Code Lack of Effect (4065)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 01/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Mw5152915 was received on (b)(6) 2024, reporting that the patient had an afx2 bifurcated stent graft implanted and on (b)(6) 2024 it was discovered that there was an indeterminate endoleak.Reportedly, the patient is being referred to a surgeon to have the afx2 bifurcated stent graft repaired.
 
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.H3 other text : device remains implanted.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.Endologix could not perform an evaluation of the device as the device remains implanted in the patient.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the indeterminate endoleak complaint is unconfirmed.This is not consistent with the reported adverse event/incident.Device, user, procedure or anatomy relatedness of this complaint could not be determined.No procedure related harms for this complaint are identified.No contributing factors were identified.The final patient status is reported as stable pending surgical consultation.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.Corrections: awareness date ¿ updated investigation finding codes - remove code 3233.Investigation conclusion codes - remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key19134661
Report Number3011063223-2024-00055
Device Sequence Number19998016
Product Code MIH
UDI-Device Identifier00818009014726
UDI-Public(01)00818009014726(17)191114
Combination Product (Y/N)N
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 04/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/14/2019
Device Model NumberBEA25-100/I16-40
Device Lot Number2011140009
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/03/2024
Supplement Date Received by Manufacturer05/22/2024
Initial Report FDA Received Date04/18/2024
Supplement Report FDA Received Date05/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/07/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age81 YR
Patient SexMale
-
-