Model Number M0061801560
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a contour vl ureteral stent was used during a stent insertion procedure performed on (b)(6) 2024.During the procedure, when the physician attempted to remove the sensor wire for stent placement, the sensor wire was stuck onto the stent.The stent became shredded, and the wire uncoiled itself.The guidewire and stent were removed from the patient without a problem.Another stent and guidewire completed the procedure.There were no patient complications reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0414 captures the reportable event of stent torn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTOUR VL
Common Device Name
STENT, URETERAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key19134820
Report Number2124215-2024-22571
Device Sequence Number4798802
Product Code FAD
UDI-Device Identifier08714729067986
UDI-Public08714729067986
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K974541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM0061801560
Device Catalogue Number180-156
Device Lot Number0032630059
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2024
Initial Report FDA Received Date04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/13/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-