| Model Number |
M0061801560 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a contour vl ureteral stent was used during a stent insertion procedure performed on (b)(6) 2024.During the procedure, when the physician attempted to remove the sensor wire for stent placement, the sensor wire was stuck onto the stent.The stent became shredded, and the wire uncoiled itself.The guidewire and stent were removed from the patient without a problem.Another stent and guidewire completed the procedure.There were no patient complications reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0414 captures the reportable event of stent torn.
|
| |
|