• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. PEEK SWIVELOCK C,4.75X19.1MM; BIO SOFT TISSUE FIXATN FASTNR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. PEEK SWIVELOCK C,4.75X19.1MM; BIO SOFT TISSUE FIXATN FASTNR Back to Search Results
Model Number PEEK SWIVELOCK C,4.75X19.1MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a shoulder arthroscopy the anchor of the device broke when putting the suture through the eyelet.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Update avoe 14-nov-2022 it was confirmed that the device didn`t break inside the patient and nothing had to be retrieved from the patient.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint is not confirmed.One unpackaged ar-2324pslc-1 suture anchor, peek-swivelock serial/batch number 14564881 was received for investigation.Visual evaluation found that the device was returned without the broken anchor for evaluation.Visual evaluation of the attached picture.The complaint cannot be confirmed based on the customer provided photo, which displays the unpackaged device with the anchor.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation, prying/leveraging the device and/or excessive force being used during insertion of the implant.No change in harm was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEEK SWIVELOCK C,4.75X19.1MM
Common Device Name
BIO SOFT TISSUE FIXATN FASTNR
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19135283
Report Number1220246-2024-02128
Device Sequence Number4658703
Product Code MAI
UDI-Device Identifier00888867098145
UDI-Public00888867098145
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K203495
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPEEK SWIVELOCK C,4.75X19.1MM
Device Catalogue NumberAR-2324PSLC-1
Device Lot Number14564881
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2022
Initial Report FDA Received Date04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-