| Model Number |
26775UF |
| Medical Device Problem Codes |
Melted (1385); Overheating of Device (1437); Use of Device Problem (1670); Dent in Material (2526)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
03/29/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.Internal karl storz reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
The incident occurred during the procedure, causing the sheath to melt and resulting in burns to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The item in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation: during the investigation of the returned article 26775uf the insulation was found to be damaged and severe thermal damage was found at the distal end.The damage indicates that too much heat has been generated on the article and the material has been burnt.It is concluded that the most probable cause for this was that the peak voltage was exceeded, or the electrode was operated for too long.The instructions for use warns not to exceed the peak voltage of max.3.0 kvp.In addition, the article is only intended to be used for short-term coagulation of minor bleeding.A longer operating time on one piece leads to a higher heat development.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|