| Model Number |
20132620-1 |
| Medical Device Problem Codes |
Off-Label Use (1494); Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/03/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that when the concerned product was used in a veterinary procedure (laparoscopic spay).It was reported that "mid surgery the light source turned off and would not turn back on.The box was very hot to touch and only displayed the red warning light once the light had already gone out.As the light source would not turn back on after trying everything they suggested we had to convert to an open surgery which is not what the client had come in for." it was confirmed that the surgery was prolonged between 15 and 60 minutes and had to be converted from laparoscopic to open surgery.Furthermore, it was confirmed that there was no injury to the patient and the patient recovered well from the surgery as expected.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in section d9 to reflect that the product was returned for evaluation.The item in question was returned to the manufacturer.The investigation is not completed yet.The investigation results are pending.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was completed on 2024-07-24.During investigation of the concerned cold light fountain xenon 100 scb several tests were performed to check the function of the device.During these tests no issues with the device were found and the xenon 100 always worked without any failures in function.The reported issue could not be duplicated.Please be aware that the reported case happened during use in a veterinary procedure, but the use of the concerned device is restricted to human medical applications.
|
| |
|