| Model Number |
ECLIPSE¿ SPEEDSCAP¿ IMPLANT SYSTEM |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on (b)(6) 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchors would not stay secure in the bone; they would pull out when tensioned on.They used another swivelock to complete the case.
|
| |
|