• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFE2000 VENTILATOR PACKAGED; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFE2000 VENTILATOR PACKAGED; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number BT-20-0002
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
Baxter received a report from a baxter technician stating the power cord had damage with exposed copper wires.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Additional Manufacturer Narrative
The baxter technician found the power cord needed to be replaced.The breathe technologies life2000® ventilation system is intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.Specifically, the system is applicable for adult patients who require the following types of ventilatory support: positive pressure ventilation, delivered invasively (via et tube) or non-invasively (via mask).Assist/control mode of ventilation.The technician replaced the power cord to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFE2000 VENTILATOR PACKAGED
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer Contact
estephania cardenas
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key19138303
Report Number1316463-2024-00047
Device Sequence Number21191631
Product Code CBK
UDI-Device Identifier00887761978201
UDI-Public00887761978201
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K170037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberBT-20-0002
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2024
Initial Report FDA Received Date04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-