| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Swelling/ Edema (4577)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Additional associated products: 42504606212 persona femur cemented plus right size 7+ lot# 64767893.42560113516 persona stem ext straight spld uncemented 16mm dia +135mm l lot# 64783575.42545001011 persona femoral central cone size small lot# 64643091.42556606205 persona femoral distal augment cemented size 7, 7+ 5mm lot# 64859431.42542006702 persona tibia fixed cemented right size d lot# 64565801.42560313512 persona stem ext 3mm offset spld uncemented 12mm dia +135mm l lot# 64675872.42522600516 persona articular surface fixed bearing (cps) right 16mm lot# 64647943.00587806532 nexgen complete knee solution trabecular metal std patella lot# 64227939.
|
| |
|
Event or Problem Description
|
|
It was reported a patient underwent right knee a two stage revision for infection.Subsequently, approximately 7 months later the patient began to have intermittent pain and swelling and 2 years five months reported an increase in pain and was prescribed physical therapy.All implants remain in place.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected.Updated: b4, b5, g3, g6, h1, h2, h3, h6, h10, h11.No products were returned, or pictures provided; visual and dimensional evaluations could not be performed.Review of the devices history records identified no deviations or anomalies during manufacturing related to the reported event.The reported products were reviewed for compatibility with no issues noted.Office notes were provided and report pain, swelling and trouble ambulating.The patient also reported fall on ice.X-rays taken during visit state no periprosthetic fracture and implants are well positioned.X-rays reviewed over two months during which shows a stable appearance of revision components.The patient reports increasing right knee pain resulting in physical therapy.Complaint is not confirmed.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Event or Problem Description
|
|
This follow-up report is being submitted to relay additional information.Upon an internal data review, it was determined that this device was reported under the incorrect manufacturer.This report should be considered void, and the event will be reported under mfr (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Upon an internal data review, it was determined that this device was reported under the incorrect manufacturer.This report should be considered void, and the event will be reported under mfr (b)(4).
|
| |
|
Search Alerts/Recalls
|