| Model Number |
M00510860 |
| Medical Device Problem Codes |
Material Deformation (2976); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0406 captures the reportable event of side car rx pushback.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a trapezoid rx was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During preparation, the trapezoid rx was inspected and observed the basket was able to open and close.However, the side car rx guidewire channel was found split into two and pushed back.Another trapezoid rx was used to complete the procedure.There were no patient complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0406 captures the reportable event of side car rx pushback.Block h11: the returned trapezoid rx was analyzed, and a product analysis found the side car rx was pushed back 2mm and torn.The reported event of side car rx pushback and torn was confirmed.Based on all available information, it is most likely that the way the device was taken out of the package is the reason why the side car rx was pushed back and torn.Therefore, the most probable root cause is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a trapezoid rx was used in the common bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2024.During preparation, the trapezoid rx was inspected and observed the basket was able to open and close.However, the side car rx guidewire channel was found split into two and pushed back.Another trapezoid rx was used to complete the procedure.There were no patient complications as a result of this event.
|
| |
|