| Catalog Number |
DWP2423 |
| Medical Device Problem Codes |
Device Difficult to Setup or Prepare (1487); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Corrections: h6 device code, d9 product available to stryker.The reported event was not confirmed, since the device was not returned for evaluation and no other evidence was provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on investigation, the root cause was attributed to a user related issue.The failure was caused by an application of excessive mechanical force during surgery.If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had a dislocation, possibly attributed to night terrors after the initial surgery.The patient underwent a revision surgery to replace the poly insert.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had a dislocation, possibly attributed to night terrors after the initial surgery.The patient underwent a revision surgery to replace the poly insert.
|
| |
|