MAUDE Adverse Event Report: TORNIER INC PERFORM REV INSERT SZ 3/4 42MM DIA +3; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC PERFORM REV INSERT SZ 3/4 42MM DIA +3; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number DWP2423
Medical Device Problem Codes Device Difficult to Setup or Prepare (1487); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Corrections: h6 device code, d9 product available to stryker.The reported event was not confirmed, since the device was not returned for evaluation and no other evidence was provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on investigation, the root cause was attributed to a user related issue.The failure was caused by an application of excessive mechanical force during surgery.If device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
It was reported that the patient had a dislocation, possibly attributed to night terrors after the initial surgery.The patient underwent a revision surgery to replace the poly insert.
 
Additional Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Event or Problem Description
It was reported that the patient had a dislocation, possibly attributed to night terrors after the initial surgery.The patient underwent a revision surgery to replace the poly insert.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORM REV INSERT SZ 3/4 42MM DIA +3
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19140370
Report Number0001649390-2024-00174
Device Sequence Number15305301
Product Code PHX
UDI-Device Identifier00846832084524
UDI-Public00846832084524
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K201315
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberDWP2423
Device Lot Number5516AY
Was Device Available for Evaluation? No
Date Returned to Manufacturer04/04/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer06/26/2024
Initial Report FDA Received Date04/18/2024
Supplement Report FDA Received Date07/18/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-