|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.New information added to the following fields: h3 and h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the operating wire coating of applicator peeled off occurred due to oxidation (rust) of the operation wire.The event can be detected and prevented by handling the device in accordance with the following instructions for use: before use, check that the hook and coil sheath are not corroded (microscopic holes, dents, discoloration, etc.) and that there is no abnormality in the operational feel.Using this product with corroded hook or coil sheath may cause the hook or coil sheath to fall off.During use, always check the endoscopic image to confirm that there is no abnormality in the tip and that there is no abnormality in the operability.In the unlikely event that the hook or coil sheath falls off during use, please retrieve it with grasping forceps.Do not squeeze, wipe, or rub the rotating clip device.It may lead to damage or deterioration of the function of the rotating clip device.Olympus will continue to monitor field performance for this device.
|