MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ORTHOMATCH CR IMPLANT IMPACTOR; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ORTHOMATCH CR IMPLANT IMPACTOR; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 71436399
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal reference number: (b)(4).H10: smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.
 
Event or Problem Description
It was reported that during a tka surgery, one (1) orthomatch cr implant impactor fractured.The procedure was resumed without any delay, using the same device.No injuries were reported as a consequence of this issue.
 
Additional Manufacturer Narrative
H6: the device was not returned for evaluation.However, the photographs were reviewed, and revealed that the body of the device is fractured.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORTHOMATCH CR IMPLANT IMPACTOR
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19142447
Report Number1020279-2024-00779
Device Sequence Number7861601
Product Code JWH
UDI-Device Identifier00885556622087
UDI-Public00885556622087
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K170648
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71436399
Device Lot Number18GSC0024
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer05/15/2024
Initial Report FDA Received Date04/19/2024
Supplement Report FDA Received Date05/16/2024
Date Device Manufactured06/19/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexFemale
-
-